Designing a Highly‑Sensitive Syphilis Antibody Lateral‑Flow Assay with Axiflow™ Components

Rapid, point‑of‑care treponemal tests are essential for interrupting Treponema pallidum transmission, yet published evaluations show that commercial assays still leave a sensitivity gap of ~1–5 percentage points and occasional hook‑effects at high antibody titres. To close that gap we engineered a new assay around Axiflow’s lateral‑flow architecture, pairing the company’s high‑performance membrane and pad portfolio with Micronic’s chimeric TP15 + TP17 + TP47 antigens.

Materials & Strip Architecture

Functional LayerAxiflow PartKey Attribute for Sensitivity
Nitrocellulose membraneAXC100‑CH25100 µm thickness, ~100 s/4 cm capillary time → optimized antigen–antibody residence
Conjugate pad16613N-T01 (Polyester)Uniform release of AuNP–antigen conjugate (6-8 OD)
Sample pad16615‑T05 (Polyester)Pre‑conditioned for serum / plasma / WB pH 7.4 ± 0.2
Blood separatorPSP60+ (Fiber blend)95 % RBC retention, <10 s separation time
Absorbent pad1222N (Cellulose)45 µL cm⁻¹ capacity prevents early flow arrest
Backing cardBPC060300 PVC300 µm rigidity for automated lamination

Performance Highlights

· Analytical LoD: 0.5 IU mL⁻¹ anti‑TP (in‑house panel, n = 25), a four‑fold improvement over a market‑leading treponemal card test run in parallel.

·Clinical sensitivity/specificity: 99.1% / 100.0% versus TPHA ± VDRL (n =400 serum & whole‑blood specimens).

Read‑time: Visible signal within 12 min; official read window 15–20 min.

·Robust matrix tolerance of PSP60+: Consistent T/C ratios with whole blood haematocrits from 35 % to 55 %.

Key Takeaways for Assay Developers

· Membrane kinetics matter: A 100-sec. membrane strikes an optimal balance between binding residency and total run‑time for IgG/IgM treponemal targets.

· Matrix‑matched pad chemistry: Pre‑treated sample and conjugate pads reduce hook effects and lot‑to‑lot drift.

· Integrated blood separation: PSP60+ enables a true whole‑blood format without external devices, widening POC usability.

·Scalable workflow: All components are supplied in reel format and are compatible with standard automated striping, drying, and lamination lines, ensuring seamless tech‑transfer into volume manufacturing.

Outcome

Leveraging Axiflow’s integrated material set enabled us to deliver a syphilis antibody LFA that meets desired regulatory & market benchmarks for treponemal diagnostics while providing faster time‑to‑result and superior analytical depth.