Axiflow Biotech – Nitrocellulose Membrane Selection Guide

1 | Core determinants

There are four key variables that dictate nitrocellulose membrane selection:
• Sample viscosity
• Desired assay time
• Analyte abundance (sensitivity goal)
• Reader / line‑quality requirements

2 | Viscosity‑first decision matrix

Sample viscosity bandTypical sample typesRecommended Axiflow grade(s)Rationale
High > 3 cPWhole blood, viscous nasal/throat swab eluate, sputum, stool slurryAXC90 or AXC100Fast flow offsets slow wetting and keeps run‑time ≤ 10 min
Moderate 1–3 cPSerum, plasma, nasopharyngeal VTM, plant sap extractsAXC100 or AXC115Balanced speed keeps test within 10–12 min, allows better LoD
Low < 1 cPSaliva, buffer‑spiked samples, purified antigen prepsAXC115, AXC130 or AXC160Slower flow increases target interaction for low‑titer analytes

3 | Expanded quick‑select table

Axiflow gradeWicking (s/4 cm)Flow descriptorIdeal sample typesSensitivity tierTypical read‑timeLine intensity
AXC90 And AXC10080 – 110Very fastUrine, Whole blood, viscous nasal swab, latex bead agglutinationGood3–6 min* (*10 mins for whole blood)Sharp
AXC115100 – 130Medium‑fastSerum, Finger‑stick blood with RBC filter,plant extracts, saliva antigen testsVery good6–15 minVery sharp
AXC130115 – 145MediumQuantitative CRP/hormone markers, fluorescent assaysHigh10–20 minVery sharp
AXC160135 – 195Slow / high‑sensitivityUltra‑low abundance analytes, veterinary assaysVery high15–30 minVery sharp

4 | Step‑by‑step selection workflow

1. Categorize your sample matrix using the viscosity bands above

2. Set maximum permitted read‑time (e.g., < 6 min for emergency POCT)

3. Define the limit of detection (LoD) in quantitative terms

4. Check detection modality & line‑width constraints (fluorescent vs. visual)

5. Prototype with at least three production lots to de-risk scale‑up

5 | Axiflow value adds

• Strong tech support for selection of best-suited membranes for your application

• 105 µm membrane on a 100 µm clear PET backing – identical stack height across all grades.

• ≤ 5 s wet‑out on all grades thanks to proprietary surfactant technology

• Custom rolls & sheets – widths 3–330 mm, lengths up to 100 meter

• QC traceability for auditors – lot‑specific wicking histogram, thickness profile, and QC certificate.

Contact our experts today at connect@axiflow.co.in, and we can help you in choosing the best-suited membranes for your application